SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 463
Example
Reset Condition
On reset, Y Max = 20 dBm; Y Scale = 7 dB.
Query
SENSe[1]|2:TRACe:Y:SCALe:PDIV?
The query returns the current scale setting of Y-axis in numerical value.
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in trigger acquisition mode, error –221 "Settings conflict"
occurs.
If limits of the values keyed in are exceeded, error –222 "Data out of range;
upper (or lower) limit exceeded; no change" occurs.
SENS:TRAC:Y:SCAL:PDIV 0.002 This command sets the Channel A Y Scale
value to 0.002 step.
SENS2:TRAC:Y:SCAL:PDIV 0.05 This command sets the Channel B Y Scale
value to 0.05 step.
SENS:TRAC:Y:SCAL:PDIV? This command queries the Channel A current
Y-axis scale setting in numerical value.