11 STATus Subsystem
474 N1911A/1912A P-Series Power Meters Programming Guide
STATus:DEVice:NTRansition is 1, then STATus:DEVice:EVENt? is set when a
power sensor is disconnected.
STATus:DEVice:PTRansition is 1, then STATus:DEVice:EVENt? is set when a
power sensor is connected.
The Channel A and B sensor error bits (3 and 4) are set to:
1, if the P-Series, N8480 Series or E-Series power sensor EEPROM has failed
or if there are power sensors connected to both the rear and front panel
connectors.
0, for every other condition.
The Front Panel key press bit (bit 14), when queried with the STATus:DEVice:EVENt?
query indicates whether any front panel keys have been pressed since power up or since
you last queried the device status register. This bit ignores the :NTRansition, and
:PTRansition registers and a :CONDition? query always returns a 0.
NOTE
Querying STATus:DEVice:EVENt? clears the STATus:DEVice:EVENt? register.