1 Power Meter Remote Operation
48 N1911A/1912A P-Series Power Meters Programming Guide
Example 6 - Making a Ratio Measurement
The following program segment can be carried out on the N1912A. It queries the
lower window to make a ratio measurement of Channel A/B. The expected power
level and resolution parameters are defaulted, leaving them at their current
settings. Some fine tuning of the measurement is carried out by setting the
averaging.
Using READ?
ABOR1
ABOR2
CONF2:POW:AC:RAT DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 512
SENS2:AVER:COUN 256
READ2:POW:AC:RAT?
READ2:POW:AC:RAT? DEF,DEF,(@2),(@1) (A second READ? query is sent to make
a Channel B -Channel A ratio measurement using fresh measurement data.)
Using INITiate and FETCh?
ABOR1
ABOR2
CONF2:POW:AC:RAT DEF,DEF,(@1),(@2)
SENS1:AVER:COUN 512
SENS2:AVER:COUN 256
INIT1:IMM
INIT2:IMM
FETC2:POW:AC:RAT?
FETC2:POW:AC:RAT? DEF,DEF,(@2),(@1) (A second FETCh? query is sent to
make a Channel B -Channel A measurement using the current measurement
data.)