TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 673
Parameters
Example
Reset Condition
On reset, the value is set to 1.
Query
TRIGger[1]|2[:SEQuence[1]|2]:COUNt?
The query returns the current setting of trigger events for a specified channel.
Query Example
Error Messages
If COUNt >1 when [SENSe[1]]|SENSe2:MRATe <character_data> is set to
NORMal or DOUBle, error –221, “Settings Conflict” occurs.
Item Description/Default Range of Values
numeric_value The number of triggered events for the measurement cycle.
DEF: the default value is 1
1 to 50
DEF
TRIG:SEQ1:COUN 10 This command sets the number of triggered
events to 10 for the Channel A measurement
cycle.
TRIG:SEQ2:COUN? This command queries the number of
triggered events for the Channel B
measurement cycle.