TRIGger Subsystem 14
N1911A/1912A P-Series Power Meters Programming Guide 679
Reset Condition
On reset, the trigger source is set to IMMediate.
Query
TRIGger[:SEQuence[1]|2]:SOURce?
The query returns the current trigger source.
Query Example
Error Messages
For dual channel power meters: if the master is changed to IMM, BUS or HOLD,
error –221 “Settings Conflict” occurs. In such situations the slave’s TRIG:SOUR
must be changed so that it is no longer a slave.
If the trigger source is changed to INT1, INT2 or EXT and SENS:SPEED has a
value of 200, error –221 “Settings Conflict” occurs.
If the trigger source is changed to INT1 or INT2 and SENS:DET:FUNC is set to
AVERage, error –221 “Settings Conflict” occurs.
If the trigger source is set to INT1 or INT2 when 8480, N8480, E4410, E9300
or E9320 (Average mode only) is connected, error –221 “Settings Conflict”
occurs.
For dual channel power meters: if the adjacent sensor is in peak mode, setting
the trigger source of 8480, N8480, E4410, E9300 or E9320 (Average mode
only) to EXTernal causes error –221 “Settings Conflict”.
TRIG:SEQ1:SOUR? This command queries the current trigger
source for Channel A.