A Calibration Factor Block Layout
762 N1911A/1912A P-Series Power Meters Programming Guide
Header Total: 6
For Each Table (tables are in the order of lower to upper):
Power, low 2 - 7.8 (signed) –127.9 to
+127.9
dBm Power for low power
flatness.
Power, high 2 - 7.8 (signed) –127.9 to
+127.9
dBm Power for high power
flatness.
Frequency (point ‘0’) 4 - 32 bit fixed 0 to Fhpb*
(2^32)
None Fhpb = freq Hz per bit
Cal factor (low power)
[a]
2 - 2.14 0.25 to 3 None Power (in watts) is divided
by this value.
Cal factor (high power)
[a]
2 - 2.14 0.25 to 3 None Power (in watts) is divided
by this value.
These table entries are repeated as shown for each frequency point
Frequency (point ‘N’) 4 - 32 bit fixed 0 to Fhpb*
(2^32)
None Fhbp = Freq Hz per bit.
Cal factor (low power) 2 - 2.14 0.25 to 3 None Power (in watts) is divided
by this value.
Cal factor (high power) 2 - 2.14 0.25 to 3 None Power (in watts) is divided
by this value.
Table size: -
See note
[a]
The table size is
dependent on the number
of frequency points.
E9300 Series Sensors:
Calibration Factor Block
Layout
No. Bytes Contents Data Format Data Range Units Notes