EasyManua.ls Logo

YASKAWA YRC1000 - 6.7.2.3 Manual Calibration Procedure

YASKAWA YRC1000
493 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Robot Settings
6.7 Home Position Calibration
6-52
HW1485509
HW1485509
6.7.2.3 Manual Calibration Procedure
Use this method if the manipulator or YRC Controller is replaced with a
new unit. This method is also used to re-enter the existing Home Position
values that appear to be correct but are not “set”.
1. Go to {MENU}
{Robot Settings} {Home Position Calibration}.
2. Locate the “ABSOLUTE DATA” white tag on the YRC Controller
containing the factory default robot encoder calibration.
Fig. 6-33: “ABSOLUTE DATA” white tag
3. Check the {Manual Entry?} checkbox.
Fig. 6-34: Manual Axis Home Position Calibration
4. Manually enter the data for each axis using the keyboard.
241 of 493

Table of Contents