R3765/67G Series Network Analyzer Operation Manual
5.8 GO/NG Measurement Using Limit Line Function
5-48
The display on the screen is as follows.
Figure 5-41 Setting of Segment 0
Set segment 1, segment 2, segment 3 and segment 4 in the same way.
SEGMENT 1 : {ADD SEGMENT}
{STIMULUS VALUE} [8] [4] [0] [MHz]
{UPPER LIMIT} [-] [4] [0] [X1]
{LOWER LIMIT} [-] [6] [5] [X1]
{Return}
SEGMENT 2 : {ADD SEGMENT}
{STIMULUS VALUE} [8] [6] [5] [MHz]
{UPPER LIMIT} [0] [X1]
{LOWER LIMIT} [-] [2] [0] [X1]
{Return}