R3765/67G Series Network Analyzer Operation Manual
8.2 Error Messages
8-3
8.2.2 Notice of Hardware Information
Ach Overload
Bch Overload
Cch Overload
(OPT11/OPT14)
Overlevel is input into A channel.
Overlevel is input into B channel.
Overlevel is input into C channel.
<How to handle>
Check the input signal level.
External Standard In.
External Trigger Ignored.
An external reference signal has been input.
An input external trigger was ignored. (That does not mean a
prohibiting state.)
<How to handle>
An external trigger (PIO-18pin) has been input in a state of not
waiting for the external trigger.
The state of waiting for the external trigger is the state of waiting
for sweep in the external trigger mode (that is, in a state that
TRIGGER[CONT] or TRIGGER[SINGLE] on the panel).
If next trigger pulse is input during a sweep in using an external
trigger source, the above error occurs.
Check the trigger setting and the specification of an external
trigger signal.