R3765/67G Series Network Analyzer Operation Manual
10.1 Preparing for a Performance Test
10-1
10 PERFORMANCE TEST
This chapter describes the test method to keep the performance of the R3765/67G series.
Contact ADVANTEST for other test methods than the items described in this chapter.
10.1 Preparing for a Performance Test
10.1.1 Warm up
Warm up the R3765/67G series for at least 30 minutes (pre-heating).
Also, warm up each calibration standards as well.
10.1.2 Preparing Measurement Instrument
The following measurement instruments are required referring to the test items listed in Table
10-1.
Table 10-1 Required Measurement Instrument for Performance Test (1 of 3)
Test items Measurement Instrument Remarks
Frequency
accuracy and
range
• Counter
Frequency : 300kHz to 3.8GHz
(R3765G Series)
300kHz to 8.0GHz
(R3767G Series)
Display : 7 digits or more
Accuracy : 0.1ppm or less
R5372 (to 18GHz)
or R5373 (to 26GHz)
(ADVANTEST-made)
Refer to
section 10.2
• RF cable
BNC-BNC, N-N Type
• 50Ω to 75Ω conversion adapter
(when OPT 12 or 13 is installed)
11852 B
(Agilent
Technologies)
Output/input
level and
flatness
• Power meter
Frequency : 300kHz to 8GHz
Power range: -21dBm to +17dBm
NRVS (R&S)
(Calibrated under the
national standard)
Refer to
section 10.3
• Power sensor
Frequency : 300kHz to 8GHz
Power range: -21dBm to +17dBm
NRV - Z51(R&S)
(DC to 18GHz)
• Network analyzer R3765/67CG
(ADVANTEST-made)
• 50Ω to 75Ω conversion adapter
(when OPT 12 or 13 is installed)
11852 B
(Agilent Technolo-
gies)
• 75Ω n-type through
(when OPT 12 or 13 is installed)
85036-60014
(Agilent Technolo-
gies)