R3765/67G Series Network Analyzer Operation Manual
7.9 Sweep
7-94
Setting sweep area
• Sweep area settings for the linear frequency sweep log frequency sweep are
performed with [START], [STOP] or [CENTER], [SPAN].
• Power sweep are performed with [START], [STOP].
• For the user frequency sweep and program sweep, set the sweep area on each
segment editing menu.
CAUTION!
If USER FREQUENCY SWEEP or PROGRAM SWEEP is set, input segments are detected
and arranged internally in increasing order of frequency.
If STOP frequency of a segment is higher than START frequency of the next segment in
the arranged segments, an error occurs.
7.9.2 Editing Segment of User Frequency Sweep
The Setting and the Explanation
Press the [MENU] to call the signal source menu. (Refer to section A.4.)
Press the {SWEEP TYPE [ ]} to call the sweep type menu.
Press the {EDIT USER SWEEP} to call the user frequency sweep segment editing
menu.
User frequency sweep segment editing menu
{SEGMENT: NUMBER} : Specifies the segment number in the range of 0 to 29.
{START} : Sets the start frequency of the specified segment.
{STOP} : Sets the stop frequency of the specified segment.
{FREQ} : Sets the frequency of the specified point when the
number of points of the specified segment is set to 1.
Reversely, if this frequency is set, then the number of
points automatically becomes 1.
{POINT} : Sets the number of points of the specified segment.
{CLEAR SEG} : Clears the specified segment.
{CLEAR ALL SEG} : Clears all segments.
CAUTION!
1. If the same segment number is edited on the segment editing menu of the program
sweep, then the user frequency segment is also changed. (Segment holds the pro-
gram sweep in common.)
2. The total number of points of each segment by user frequency sweep cannot exceed
1201 points. (The maximum number of measurement points is 1201 points.)
1
2
3
4

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3765G Series and is the answer not in the manual?

ADVANTEST R3765G Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3765G Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals