R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-42
7.5.6 Interpolating Calibration Data
When {INTERPOLATE ON/OFF} is set to ON, the calibration data is performed the interpolation
error correction measurement even if the stimulus setting shown below is changed during error
correction measurement (during calibration).
Change of the sweep range (within the calibration range)
Change of the sweep type (within the constraints)
Change of the sweep point number
Table 7-1 Combination of Interpolatable Sweep Type (o; possible, x; impossible)
The status display beside the scale showing the calibration status shows the following meaning.
Table 7-2 Status Display
Normalize
Normalize
&
Isolation
1-port
calibration
2-port
calibration
3-port
calibration
4-port
calibration
Normal correction
Interpolative correction
Abnormal correction
Cor
C?
C!
Cor
C?
C!
Cor
C?
C!
C2
C2?
C2!
C3
C3?
C3!
C4
C4?
C4!
Sweep type under
calibration
The sweep type
at the present
Linear
Sweep
Log
Sweep
User
Sweep
Program
Sweep
Power
Sweep
Linear Sweep
Log Sweep
User Sweep
Program Sweep
Power Sweep
*1 *2
*1: Obtain the calibration data (1 point) corresponding to CW frequency from the linear sweep
frequency range at the calibration and make all the points to the same calibration data.
*2: Compensate with the output level only when CW frequency is the same.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3765G Series and is the answer not in the manual?

ADVANTEST R3765G Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3765G Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals