R3765/67G Series Network Analyzer Operation Manual
10.2 Frequency Accuracy and Range
10-4
10.2 Frequency Accuracy and Range
Testing procedure
Setup the R3765/67G series as follows:
For the AG type, connect the other side of the cable to the SOURCE terminal.
For the BG, the CG type or OPT11/13/14, connect the other side of the cable to TEST
PORT1.
Figure 10-1 Frequency Accuracy and Range
Set the R3765/67G series as follows:
Span : 0Hz
Trigger mode : HOLD
Change any center frequency in the range of 300kHz to 3.8GHz (R3767G Series; to
8.0GHz).
<Check> : Counter read frequency < center frequency ± center frequency × 10 × 10
-6
(Example)
When the center frequency is 100MHz : 100MHz ± 1kHz (in 23°C to ±5°C)
That is, 99999000Hz to 100001000Hz is enable.
1
TEST PORT 1 TEST PORT 2
* Connect to SOURCE terminal when AG type.
To counter input
RF cable
* For OPT12 or 13, connect the 50Ω to 75Ω conversion adapter to TEST PORT 1.
2
3
4