EasyManuals Logo

ADVANTEST R3765G Series User Manual

Default Icon
446 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #173 background imageLoading...
Page #173 background image
R3765/67G Series Network Analyzer Operation Manual
7.4 RESPONSE Block
7-29
Depending on the measurement frequency, the IF RBW
AUTO function automatically switches the resolution
bandwidth during sweeping since the resolution
bandwidth cannot be distinguished within the sweep
frequency range.
When [AUTO] is set, the resolution bandwidth is
automatically selected according to the measurement
frequency range as shown in the table below. Then the
measurement is performed.
300Hz 3.4msec/POINT
200Hz 5.0msec/POINT
150Hz 7.0msec/POINT
100Hz 11.0msec/POINT
70Hz 14.0msec/POINT
50Hz 19.0msec/POINT
40Hz 26.1msec/POINT
30Hz 34.9msec/POINT
20Hz 50.1msec/POINT
15Hz 70.1msec/POINT
10Hz 99.3msec/POINT
Measurement frequency Resolution bandwidth
300kHz to 450kHz 10kHz
450kHz to 700kHz 15kHz
700kHz to 8GHz 20kHz
Resolution bandwidth Maximum sweeping per point