R3765/67G Series Network Analyzer Operation Manual
7.4 RESPONSE Block
7-29
Depending on the measurement frequency, the IF RBW
AUTO function automatically switches the resolution
bandwidth during sweeping since the resolution
bandwidth cannot be distinguished within the sweep
frequency range.
When [AUTO] is set, the resolution bandwidth is
automatically selected according to the measurement
frequency range as shown in the table below. Then the
measurement is performed.
300Hz 3.4msec/POINT
200Hz 5.0msec/POINT
150Hz 7.0msec/POINT
100Hz 11.0msec/POINT
70Hz 14.0msec/POINT
50Hz 19.0msec/POINT
40Hz 26.1msec/POINT
30Hz 34.9msec/POINT
20Hz 50.1msec/POINT
15Hz 70.1msec/POINT
10Hz 99.3msec/POINT
Measurement frequency Resolution bandwidth
300kHz to 450kHz 10kHz
450kHz to 700kHz 15kHz
700kHz to 8GHz 20kHz
Resolution bandwidth Maximum sweeping per point