R3765/67G Series Network Analyzer Operation Manual
7.5 Calibration
7-37
(3) Normalize & isolation calibration
The Setting and the Explanation
Set the instrument to the transmission measurement mode.
Press [CAL] to call the calibration menu (1/2).
Press {CAL MENU} to call the calibration selection menu.
Press {2 PORT CAL MENU} to call the 2-port calibration selection menu.
Press {NORMALIZE & ISON} to call the normalization and isolation calibration menu.
Connect the through standard between the test ports and press {THRU}.
The message Wait for Sweep . is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Connect the load standard to each test port and press {ISOLATION}.
The message Wait for Sweep . is displayed and the calibration data is obtained. The
calibration is completed when the message disappears. (Note)
Press {DONE NORM & ISO} to complete the calibration of the normalize & isolation
calibration.
Connect DUT to perform the measurement.
Note: Do not move the R3765/67G series, the cable, the connector, the standard, and
the others when the message Wait for Sweep . is displayed.
When the set condition is changed during the message is displayed, the
message Calibration aborted . is displayed and the current calibration data
cannot be acquired.
1
2
3
4
5
6
7
8
9

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ADVANTEST R3765G Series and is the answer not in the manual?

ADVANTEST R3765G Series Specifications

General IconGeneral
BrandADVANTEST
ModelR3765G Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals