EasyManuals Logo

ADVANTEST R3765G Series User Manual

Default Icon
446 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #349 background imageLoading...
Page #349 background image
R3765/67G Series Network Analyzer Operation Manual
10.7 Noise Level
10-23
• Measure the noise level at Input C.
(a) [MEAS] → {TEST-PORT CONECTION} → {S31 (P1 → P3)}
(b) [SYSTEM] → {SERVICE MENU} → {SERVICE MODE} → {MAINTENANCE
MEAS MENU} → {C}
(c) [AVG] → {SMOOTHING ON} → {SMOOTHING APERTURE} → [2] → [0] → [x1].
(d) [SYSTEM] → [SYSTEM] → {SERVICE MENU} → {SERVICE MODES}
→ {SOURCE PLL OFF}.
(e) Read the noise level indicated by the marker.
<Check> : Noise level (3kHz bandwidth)
OPT 11
300kHz to 40MHz : -73dB or less
40MHz to 3.8GHz : -78dB or less
3.8GHz to 8.0GHz : -68dB or less (R3767CG only)
OPT 13
300kHz to 40MHz : -65dB or less
40MHz to 3.8GHz : -70dB or less
OPT14
300kHz to 40MHz : -73dB or less
40MHz to 3.8GHz : -73dB or less
3.8GHz to 8.0GHz : -63dB or less (R3767CG only)

Table of Contents