CALibration Subsystem 4
N1911A/1912A P-Series Power Meters Programming Guide 219
Syntax
Example
Reset Condition
On reset, automatic calibration is disabled.
Query
CALibration[1]|2:AUTO?
The query always returns a value of 0.
Error Messages
– If this command is set to ON and an 8480 Series, E-Series, N8480 or P-Series
power sensor is connected the error –241, “Hardware missing” occurs.
– If the calibration was not carried out successfully the error –231, “Data
Questionable; CAL ERROR” occurs. If you are using an N1912A the error
message specifies which channel failed calibration.
– If there is no sensor connected, the error –241, “Hardware Missing” occurs.
CAL1:AUTO ONCE This command causes the power meter to
perform a calibration on Channel A.
:AUTO
0|OFF
1|ON
Space
?
CAL
1
2
ONCE
If the command is set to ON when a N1920 is connected, auto cal is enabled.