11 STATus Subsystem
470 N1911A/1912A P-Series Power Meters Programming Guide
Parameters
Query
:ENABle?
The query returns a 15 bit decimal-weighted number representing the contents of
the Enable Register of the SCPI Register Set being queried. The format of the
return is <NR1> in the range of 0 to 32767 (2
15
–1).
:NTRansition <NRf>|<non-decimal numeric>
This command sets the Negative Transition Register of the SCPI Register Set you
require to control. The parameter value, when rounded to an integer and
expressed in base 2 has its first 15 bits written into the Negative Transition
Register of the SCPI Register Set concerned. The last bit (bit 15) is always set to 0.
Type Description Range of Values
NRf The value used to set the Enable Register.
0 to 2
16
–1
non-decimal numeric