Remote Control Commands
R&S
®
SMW200A
1226User Manual 1175.6632.02 ─ 30
Setting parameters:
<Pathname> string
Example:
SYSTem:RCL "/var/user/temp/Test"
Loads the file Test.savrcltxt from the direc-
tory /var/user/temp/.
Usage: Setting only
:SYSTem:PROTect<ch>[:STATe] <State>[, <Key>]
Activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
See also "Protection" on page 752
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 1)
Setting parameters:
<Key> integer
The respective functions are disabled when the protection level
is activated. No password is required for activation of a level.
A password must be entered to deactivate the protection level.
The default password for the first level is 123456.
This protection level is required to unlock internal adjustments
for example.
Example: To activate protection level:
SYSTem:PROTect1:STATe 1
Internal adjustments or hostname cannot be changed.
To unlock protection level 1:
SYSTem:PROTect1:STATe 0,123456
Internal adjustments are accessible.
Manual operation: See "Protection Level/Password" on page 754
:SYSTem:SECurity:VOLMode[:STATe] <SecPassWord>, <MmemProtState>
Activates volatile mode, so that no user data can be written to the internal memory per-
manently.
To enable volatile mode, reboot the instrument. Otherwise the change has no effect.
Parameters:
<MmemProtState> 0 | 1 | OFF | ON
*RST: 0
SYSTem Subsystem