Remote Control Commands
R&S
®
SMW200A
1203User Manual 1175.6632.02 ─ 30
1
User-defined adjustment value.
The instrument is no longer in the calibrated state.
The calibration value is, however, not changed. The instrument
resumes the calibrated state if you send
SOURce:ROSCillator:INTernal:ADJust:STATe 0.
*RST: n.a. (factory preset: 0)
Example: See Example "Configuring the reference oscillator"
on page 1199.
Manual operation: See "Adjustment Active" on page 478
14.19.18 RF Ports Alignment
Option: R&S SMW-B9 and R&S SMW-K545
See also Chapter 14.18.5, "RF Ports Alignment", on page 935.
:SOURce<hw>:RFALignment:CALibrated:FREQuency?...................................................1203
:SOURce<hw>:RFALignment:CALibrated:POWer:PEP?.................................................. 1203
:SOURce<hw>:RFALignment:FOFFset?.........................................................................1204
:SOURce<hw>:RFALignment:DATTenuation?................................................................. 1204
:SOURce<hw>:RFALignment:CORRection:PHASe?........................................................1204
:SOURce<hw>:RFALignment:CORRection:LEVel?..........................................................1204
:SOURce<hw>:RFALignment:CORRection:IQDelay?.......................................................1205
:SOURce<hw>:RFALignment:CALibrated:FREQuency?
Queries the frequency for that the calibration data is valid.
Return values:
<CalibratedFreq> float
Increment: 0.01
Example: See Example "Retrieving details on the calibrated values"
on page 935.
Usage: Query only
Manual operation: See "Calibrated Frequency" on page 659
:SOURce<hw>:RFALignment:CALibrated:POWer:PEP?
Queries the PEP for that the calibration data is valid.
Return values:
<PEP> float
Increment: 0.01
Example: See Example "Retrieving details on the calibrated values"
on page 935.
SOURce Subsystem