Signal Routing and System Configuration
R&S
®
SMW200A
140User Manual 1175.6632.02 ─ 30
Connector Direction/
Color
Enabled in the "Fading and Base-
band Configuration"
Enabled in the "I/Q
Stream Mapping"
I/Q Yes Yes
Yes Not used
RF Not used Yes
Not used Not used
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:DIRection? on page 925
:SCONfiguration:EXTernal:FADer<ch>:DIRection? on page 925
:SCONfiguration:EXTernal:BBMM<ch>:DIRection? on page 925
:SCONfiguration:EXTernal:IQOutput<ch>:DIRection? on page 925
:SCONfiguration:EXTernal:RF<ch>:DIRection? on page 925
External Instrument
Accesses a dialog with settings that help you to establish the connection to the exter-
nal instruments, see Chapter 4.4.3.2, "External Instrument Configuration Settings",
on page 143.
Note:
Each external instrument must be connected via LAN or USB to the R&S SMW.
In most test setups, all instruments must use a common reference frequency signal,
either from an external common reference source or using the distributed reference
frequency of the R&S SMW.
See also:
Chapter 4.9.5, "How to Cable the Instruments in MIMO Test Setups", on page 194
Chapter 4.9.6, "How to Connect and Configure External Instruments", on page 195.
I/Q Connection
Depends on the interface type.
Analog I/Q interfaces:
Selects the type of the I/Q output signal
Digital I/Q interfaces:
Indicates the status of the I/Q connection of the digital interfaces.
The indication includes a status LED and an icon, where the icon depends on the
connector (DIG I/Q or HS DIG I/Q) used.
System Configuration Settings