Getting Started
R&S
®
SMW200A
30User Manual 1175.6632.02 ─ 30
4. Tighten all screws at the rack brackets with a tightening torque of 1.2 Nm to secure
the R&S SMW in the rack.
To unmount the R&S SMW from a rack
1. Loosen the screws at the rack brackets.
2.
WARNING! The R&S SMW can be heavy, e.g., if fully equipped. Use a lifting
equipment, see also "Lifting and carrying the product" on page 19.
Bring the lifting equipment to shelf height.
3. Remove the R&S SMW from the rack.
4. If placing the R&S SMW on a bench top again, unmount the adapter kit from the
R&S SMW. Follow the instructions provided with the adapter kit.
3.1.5 Considerations for Test Setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
Use high-quality shielded cables, especially for the following connector types:
BNC
Double-shielded BNC cables.
How to: "To connect to non-screwable connectors (BNC)" on page 33
SMA
Double-shielded SMA cables.
How to: "To connect to REF IN/REF OUT (reference = 1 GHz)" on page 35
USB
Double-shielded USB cables.
How to: Chapter 3.1.8, "Connecting USB Devices", on page 32.
See also Chapter 15.6, "Measuring USB Cable Quality", on page 1262.
LAN
At least CAT6 STP cables.
How to: Chapter 3.1.7, "Connecting to LAN", on page 31
Always terminate open cable ends.
Ensure that connected external devices comply with EMC regulations.
Use cables of the same type and equal length for connection to the I/Q and I/Q Bar
interfaces of the instrument.
How to: See "To connect instruments to the I/Q analog interface" on page 196.
Use the cable R&S SMU-Z6 for connection to the DIG I/Q interfaces of the instru-
ment. The cable is available under material number 1415.0201.02.
How to: "To connect instruments to the I/Q digital interface" on page 198.
Use the cable R&S DIGIQ-HS for connection to the HS DIG I/Q interfaces of the
instrument. The cable is available under material number 3641.2948.03.