EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - AWGN Test Signal

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Internal Baseband Source
R&S
®
SMW200A
349User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:BB:DVB:NOTCh<ch>:BWIDth[:ABSolute] on page 1060
[:SOURce<hw>]:BB:EUTRa:NOTCh<ch>:BWIDth[:ABSolute] on page 1060
[:SOURce<hw>]:BB:OFDM:NOTCh<ch>:BWIDth[:ABSolute] on page 1060
[:SOURce<hw>]:BB:ONEWeb:NOTCh<ch>:BWIDth[:ABSolute] on page 1060
Notch Bandwidth /% ← Notch definition table
Sets the notch bandwidth relative to Clock Frequency.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:NOTCh<ch>:BWIDth:RELative on page 1061
etc. for the corresponding option:
[:SOURce<hw>]:BB:DVB:NOTCh<ch>:BWIDth:RELative on page 1061
[:SOURce<hw>]:BB:EUTRa:NOTCh<ch>:BWIDth:RELative on page 1061
[:SOURce<hw>]:BB:OFDM:NOTCh<ch>:BWIDth:RELative on page 1061
[:SOURce<hw>]:BB:ONEWeb:NOTCh<ch>:BWIDth:RELative on page 1061
Apply
Select "Apply" to adopt the configured settings.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:NOTCh:APPLy on page 1061
etc. for the corresponding option:
[:SOURce<hw>]:BB:DVB:NOTCh:APPLy on page 1061
[:SOURce<hw>]:BB:EUTRa:NOTCh:APPLy on page 1061
[:SOURce<hw>]:BB:OFDM:NOTCh:APPLy on page 1061
[:SOURce<hw>]:BB:ONEWeb:NOTCh:APPLy on page 1061
5.8.4 AWGN Test Signal
Option: R&S SMW-K811.
See also Chapter 5.8.1, "Required Options", on page 345.
Access:
Select "Baseband > ARB > General > Test Signal Form > AWGN".
The remote commands required to define these settings are described in Chap-
ter 14.19.4.3, "SOURce:BB:ARBitrary Subsystem", on page 1019.
To generate an AWGN test signal
1. Select "Test Signal Form > AWGN".
2. Click "Create Test Signal".
The dialog "AWGN Test SIgnal" opens. Optionally, adjust the number of samples of
the test signal.
Generating Notched Signals

Table of Contents

Related product manuals