Configuring the RF Signal
R&S
®
SMW200A
562User Manual 1175.6632.02 ─ 30
8.8.5.2 How to Generate a Pulse Modulated Signal
The following example uses the internal pulse generator.
To set the frequency and level of the RF signal
1. Press [PRESET].
2. In the status bar, set "Frequency = 4 GHz".
3. Set "Level = -25 dBm".
To configure the pulse generator
1. Select "RF" > "Analog Modulation" > "Pulse Generator".
2. Select "Pulse Mode = Double".
3. Set "Pulse Period = 10 us".
4. Set "Pulse Width = 2 us".
5. Set "Double Pulse Width = 1.2 us".
6. Set "Double Pulse Delay = 4.5 us".
7. Activate "Pulse Output State".
8. Set "Trigger Mode = Auto".
In the "Pulse Graph" tab, you can view the generated pulse signal graphically.
To enable pulse modulation
1. Select the "Pulse Modulation" tab.
2. Set "State = I" to activate pulse modulation.
3. Set "RF > On" to enable signal output.
8.9 Improving Level Performance
To adjust the RF output signal to specific needs in your application optimally, the
R&S SMW provides different functions:
Attenuator
The R&S SMW is equipped with a step attenuator that enables you to vary the
amplitude of the RF signal in a wide range. It is characterized by low VSWR (volt-
age standing wave ratio) over the full level and frequency range, and provides
highest level accuracy and noise suppression.
See Chapter 8.9.1, "Attenuator", on page 563.
Automatic Level Control (ALC)
The automatic level control system ensures stable RF signals with highest absolute
level accuracy over temperature and time.
See Chapter 8.9.2, "ALC - Automatic Level Control", on page 566.