EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Source:sweep Subsystem

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1205User Manual 1175.6632.02 ─ 30
Return values:
<Level> float
Increment: 0.01
Example: See Example "Retrieving details on the calibrated values"
on page 935.
Usage: Query only
Manual operation: See "Level" on page 660
:SOURce<hw>:RFALignment:CORRection:IQDelay?
Queries the I/Q delay applied for compensation of the frequency response of the sig-
nal.
Return values:
<IQdelay> float
Increment: 1E-12
Example: See Example "Retrieving details on the calibrated values"
on page 935.
Usage: Query only
Manual operation: See "I/Q Delay" on page 660
14.19.19 SOURce:SWEep Subsystem
The SOURce:SWEep subsystem contains the commands for configuring RF sweep sig-
nals.
The keyword [:FREQuency] can be omitted, then the commands are SCPI-com-
pliant.
To activate an RF sweep mode, use the following commands:
RF frequency sweep: SOURce:FREQuency:MODE SWEep
(SOURce:FREQuency:MODE CW (off))
RF level sweep:SOURce:POWer:MODE SWEep (SOURce:POWer:MODE CW
(off))
RF combined sweep: SOURce:FREQuency:MODE SWEep
(SOURce:FREQuency:MODE CW (off)
All sweeps, including the LF sweep, can be set independently of each other.
All sweeps can be set independently of each other.
See Chapter 8.7.1, "Signal Generation and Triggering in the Sweep and List Modes",
on page 493.
SOURce Subsystem