EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - External RF and I;Q Instruments

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
921User Manual 1175.6632.02 ─ 30
:SCONfiguration:OUTPut:MAPPing:STReam<st>:FOFFset <SmFreqOffset>
Option: R&S SMW-B9 - supported if "Signal Outputs = Analog Only"
Sets an absolute frequency offset.
Thus, the streams routed to the RF, I/Q Out and BBMM connectors are shifted in the
frequency domain.
Parameters:
<SmFreqOffset> float
Range: depends on the installed options, e.g. -60E6 to
+60E6 (base unit)
Increment: 0.01
*RST: 0
Value range depends also on the system confugration mode and
signal outputs. For details, see data sheet.
Example: See Example "Distributing the I/Q streams to the output connec-
tors" on page 904.
Manual operation: See "f
offset
, MHz" on page 133
:SCONfiguration:OUTPut:MAPPing:STReam<st>:POFFset <SmPhasOffset>
Option: R&S SMW-B9 - supported if "Signal Outputs = Analog Only"
Sets the phase offset of the corresponding stream.
Parameters:
<SmPhasOffset> float
Range: -999.99 to 999.99
Increment: 0.01
*RST: 0
Example: See Example "Distributing the I/Q streams to the output connec-
tors" on page 904.
Manual operation: See "Phase Offset" on page 134
14.18.4 External RF and I/Q Instruments
The following commands configure external RF and I/Q instruments.
SCONfiguration Subsystem