Network Operation and Remote Control
R&S
®
SMW200A
817User Manual 1175.6632.02 ─ 30
3. Execute the following commands on the controller:
a) Open the port to the instrument.
CALL IBFIND("DEV1", generator%)
b) Inform the controller about instrument address.
CALL IBPAD(generator%, 28)
c) Reset the instrument.
CALL IBWRT(generator%, "*RST;*CLS")
d) Set the instrument to new address.
CALL IBWRT(generator%, "SYST:COMM:GPIB:ADDR 18")
e) Inform the controller about new address.
CALL IBPAD(generator%, 18)
The GPIB address of the instrument is changed.
4. To return to manual operation, press the LOCAL key at the front panel.
13.7.7 Setting Up a Remote Control Connection over USB
For remote control via the USB connection, the PC and the instrument must be con-
nected via the USB type B interface. A USB connection requires the VISA library to be
installed. VISA detects and configures the R&S instrument automatically when the
USB connection is established. You do not have to enter an address string or install a
separate driver.
Starting a remote control session over USB
1. Connect instrument and controller using USB cable. Switch them on.
2. Execute the following commands on the controller:
a) Open the port to the instrument.
viOpen (..., "USB::0x0AAD::0x0092::100001::INSTR", ...)
b) Reset the instrument.
viRST (generator%, "*RST;*CLS")
c) Set 2 GHz frequency
viPrintf (..., "SOUR:FREQ 2GHz\n")
d) Set 20 dBm output level
viPrintf (..., "SOUR:POW -20dBm\n")
The RF frequency and signal level of the instrument are changed.
3. To return to manual operation, press the [LOCAL] key.
13.7.8 How to Trace Messages with the LXI Web Browser Interface
The remote trace functionality allows you to trace commands and messages
exchanged via a remote control interface of the R&S SMW.
Controlling the R&S
SMW Remotely