Signal Routing and System Configuration
R&S
®
SMW200A
138User Manual 1175.6632.02 ─ 30
Remote connection status can be toggled with the parameter Remote Connection Sta-
tus.
Note: Disconnecting the remote connection to SZU presets several RF parameters of
the R&S SMW to their default value.
This preset applies in particular to the RF frequency and RF level; the RF output is
deactivated.
Note:
External instruments connected to the R&S SMW are preset (:DEVice:PRESet)
whenever:
●
The connection to the external instrument is established or terminated ("Connect/
Disconnect All Remote" or "Rem. Conn > On/Off")
●
The R&S SMW is preset ([PRESET]).
To ensure that the remote connections to the external instruments will be automatically
established after instrument's start-up or a reboot, set Auto Connect > "On".
See also:
●
Chapter 4.9.5, "How to Cable the Instruments in MIMO Test Setups", on page 194
●
Chapter 4.9.6, "How to Connect and Configure External Instruments", on page 195.
Remote command:
:SCONfiguration:EXTernal:REMote:CONNect[:ALL] on page 924
:SCONfiguration:EXTernal:REMote:DISConnect[:ALL] on page 924
Preset behavior: Keep connections to external instruments
Sets the effect of R&S SMW preset ([PRESET] or *RST) on the following:
●
The connection status between the instrument and the connected external instru-
ments.
●
The R&S SMW itself and the preset values.
●
The connected instruments.
"Preset behavior..." Off On
R&S SMW and preset values Standard R&S SMW preset
See Table 11-1.
Depends on the connected external instrument:
●
R&S SZU
RF = 60 GHz
RF level = -30 dB
RF > Off
●
Otherwise, standard R&S SMW preset.
For example, RF = 1 GHz
Excluded from preset:
●
"I/Q Output Type"
See also
Connection status Terminated ("Disconnect All Remote") Unchanged
Connected instruments External instrument preset (:DEVice:PRESet) External instrument preset
Remote command:
:SCONfiguration:EXTernal:PBEHaviour on page 924
Connector Name
Displays the connector name.
System Configuration Settings