File and Data Management
R&S
®
SMW200A
678User Manual 1175.6632.02 ─ 30
signal relevant parameters of the instrument, the "Factory Preset" affects almost all
instrument settings.
For information on the default values of further parameters, refer to the description of
the corresponding remote commands.
Table 11-1: Key parameters affected by preset and factory preset
Parameter Preset value Preset Factory
Preset
RF frequency 1 GHz x x
RF level (RF output) off x x
RF OFF mode - - x
Offsets 0 x x
Modulation state off x x
Uninterrupted level settings off x x
Level attenuator mode auto x x
Level ALC (internal level control) auto x x
Level UCOR (user correction) off x x
LF output state off x x
Sweep state off x x
List mode state off x x
Reference frequency settings (reference oscillator) - - x
Power on settings (Level/EMF) - - x
Network settings - - x
Hostname
1)
- - x
GPIB address - - x
Start/Stop display update - - x
Display and keyboard settings - - x
Preset behavior: Keep connections to external instru-
ments
off - x
Password and settings protected by passwords (e.g. dis-
abled LAN or USB)
- - -
Security settings - - -
User files (setups
2)
, data lists, etc.)
- - -
●
1)
Hostname syntax rs<inst><Serial Number> has changed to
<INST>-<Serial Number> (FW 3.20.200.xx and later)
●
2)
UserPreset.savrcltxt is renamed as UserPresetInactive.savrcltxt;
an existing file with the same name is overwritten.
Restoring the (Default) Instrument Configuration