General Instrument Functions
R&S
®
SMW200A
728User Manual 1175.6632.02 ─ 30
The signal can serve as a common external trigger signal for triggering of several
R&S SMW, see Example "Triggering several R&S SMW instruments simultaneously"
on page 649.
Remote command:
:OUTPut:USER<ch>:TRIGger[:IMMediate] on page 951
Threshold USER1-3 Input
Sets the high/low threshold in volts for the signal at the USER1-3 connectors.
The input signal is defined with the parameter Signal. The same threshold applies for
all 3 connectors and any input signal.
Remote command:
[:SOURce]:INPut:USER:TRIGger:LEVel on page 951
[:SOURce]:INPut:USER:CLOCk:LEVel on page 951
Threshold USER4-6 Input/Threshold Pulse Input
Sets the high/low threshold in volts for the signal at the USER4-6 connectors.
The input signal is defined with the parameter Signal. The same threshold applies for
all 3 connectors and any input signal (e.g. external pulse modulation or trigger signal).
Remote command:
[:SOURce]:INPut:USER:PULM:LEVel on page 951
Impedance Clock/Trigger Input
Selects the input impedance for the external trigger/clock inputs.
Set the value to 1 kΩ/GND for high clock rates.
Remote command:
[:SOURce]:INPut:USER:TRIGger:IMPedance on page 952
[:SOURce]:INPut:USER:CLOCk:IMPedance on page 952
Clock Input Slope
Sets the polarity of the active slope of an externally applied clock signal.
Remote command:
[:SOURce]:INPut:USER:CLOCk:SLOPe on page 952
Trigger Input Slope
Sets the polarity of the active slope of an applied instrument trigger.
Remote command:
[:SOURce]:INPut:USER:TRIGger:SLOPe on page 952
12.2.6 Local Connectors Settings
Access:
1. Perform one of the following:
● In the "Block Diagram", select the T/M/C status LEDs to access the "Trigger
Marker Clock" dialog and:
Configuring Local and Global Connectors