Signal Routing and System Configuration
R&S
®
SMW200A
166User Manual 1175.6632.02 ─ 30
"Overflow Hold"
Indicates an overload since the last reset for evaluating the measure-
ment.
To reset the state, use one of the following:
●
Select "Reset"
●
Start a new measurement with the "Auto Level Set" function
●
Change the value of one of the monitored parameters.
Remote command:
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe? on page 963
"Reset"
Resets the "Overload Hold" indication.
Remote command:
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe? on page 963
4.6.2 Baseband Input Settings in the Wideband Baseband Generator
Option: R&S SMW-B9
To access and configure the "Baseband Input" settings
The "BB Input" block provides access to the settings for frequency offset and path
gain, and to the available configuration parameters of the external signal.
The digital baseband input interfaces HS DIG I/Q are located on the R&S SMW rear
panel, see Chapter 4.3.1, "Physical Location of the Input and Output Interfaces",
on page 114.
In the default instrument state, the block diagram displays the connector in an unfolded
state. The digital input interfaces are inactive (gray LEDs).
1. To display the "BB Input" block if the connectors are folded, select the symbol
"CODER 1/2"
1
2
Figure 4-12: Representation of the digital baseband input interface HS DIG I/Q
1 = Digital input interface state: gray = inactive; blue = active
2 = Connector name
2. To access the dialog box for configuring the external signal input depending on the
current "System Configuration > Fading/Baseband Configuration", select "BB Input
> Baseband Input Settings".
Digital Baseband Input Settings