Remote Control Commands
R&S
®
SMW200A
888User Manual 1175.6632.02 ─ 30
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped.
To define the output state, use the command :OUTPut<hw>[:STATe].
Example:
:OUTPut1:PROT:CLE
Resets the protective circuit of the RF output.
Usage: Event
Manual operation: See "Overload" on page 566
:OUTPut<hw>:PROTection:STATe <State>
Attenuates the RF output signal for about 40 dB to protect external devices against
internal signals.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
:OUTPut1:PROT:STAT ON
Attenuates the RF Output signal for about 40 dB.
Manual operation: See "Overload" on page 566
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
Return values:
<Tripped> 0 | 1 | OFF | ON
*RST: 0
Example:
:OUTPut1:PROT:TRIP
queries the state of the protective circuit of the RF output.
Response: 0
the protective circuit has not tripped.
Usage: Query only
Manual operation: See "Overload" on page 566