Minimum Locking Range
Indicates the minimum locking range for the selected External Reference Frequency
and the Synchronization Bandwidth.
The settings depend on the versions of the RF hardware, see Appendix: RF Hardware
Versions, on page 1289.
Remote command:
[:SOURce]:ROSCillator:EXTernal:MLRange? on page 1201
8.5.3 Reference Output Settings
Access:
1. Select "RF" > "RF Frequency" > "Reference Frequency".
2. Select "Reference Output".
In the "Reference Output" tab, you can set the reference frequency value at the
output connectors.
As a result of parameter dependencies, "Preset This Parameter" sometimes does
not affect output dialogs.
The remote commands required to define these settings are described in Chap-
ter 14.19.17, "SOURce:ROSCillator Subsystem", on page 1198.
Settings:
Reference Output/1 GHz Reference Output............................................................... 477
Show Connector..........................................................................................................478
Reference Output/1 GHz Reference Output
Selects frequency reference output signal for downstream instruments.
The settings depend on the versions of the RF hardware, see Appendix: RF Hardware
Versions, on page 1289.
"Off"
Deactivates the reference signal output.
"10 MHz"
Derives a signal with 10 MHz frequency from the internal reference
oscillator and provides this signal at the output.
"100 MHz"
Option: R&S SMW-K704
Derives a signal with 100 MHz frequency from the internal reference
oscillator and provides this signal at the output.
"1 GHz"
Option: R&S SMW-K703
Provides the 1 GHz reference frequency signal at the output.