Signal Routing and System Configuration
R&S
®
SMW200A
179User Manual 1175.6632.02 ─ 30
●
If Common to all Channels > "Off", you can set the sample rate per channel individ-
ually.
●
If Common to all Channels > "On", the value resembles the value set with the
parameter Value.
For information on the max. sample rate and the aggregated sample rate, depending
on the number of active channels, observe the indication Aggregated Link Sample
Rate.
Tip: For tests at max. sample rate, set the sample rate per channel to the max. value
and activate the channels one by one subsequently.
See also Chapter 4.3.4, "Supported Digital Interfaces Depending on the System Con-
figuration", on page 119.
Remote command:
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:CHANnel<st0>:SRATe
on page 1151
Active
Activates the channel.
Remote command:
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:CHANnel<st0>:STATe
on page 1151
Aggregated Link Sample Rate
Indicates the aggregated sample rate and the maximum sample rate, where:
●
The aggregated sample rate is the sum of the sample rates of all active channels.
It cannot exceed the maximum sample rate.
●
The R&S SMW supports two standard max. sample rate values.
This parameter indicates the currently selected one, where the max. sample rate is
selected depending on the capabilities of the transmitter/receiver.
For more information, see data sheet.
Remote command:
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:SRATe:SUM? on page 1152
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:SRATe:MAX? on page 1152
4.7.3 Signal Output Settings
Access:
► Select "I/Q Digital > I/Q Digital Settings > Signal Output".
"User Variation" and "Group Delay Compensation" settings are not available, if:
● R&S SGT is connected to the DIG I/Q interface of the R&S SMW
● "Signal Outputs > Digital Only (HS)"
I/Q Digital Output Settings