Remote Control Commands
R&S
®
SMW200A
863User Manual 1175.6632.02 ─ 30
<file_name> String parameter to specify the file name with extension
*.savrcltxt.
Example: See Example "Saving and loading current settings"
on page 853.
Usage: Event
Manual operation: See "Save" on page 683
:MEMory:HFRee?
Returns the used and available memory in Kb.
Return values:
<TotalPhysMemKb> integer
Total physical memory.
<ApplicMemKb> integer
Application memory.
<HeapUsedKb> integer
Used heap memory.
<HeapAvailableKb> integer
Available heap memory.
Usage: Query only
14.8 CALibration Subsystem
The CALibration subsystem contains the commands needed for performing internal
adjustments. This procedure is triggered by the query commands.
Suffix <hw>
Suffix Value range Description
CALibration<hw>
[1] to 2
CALibration[1] = Path A
(optional suffix)
CALibration2 = Path B
(mandatory suffix)
Understanding the query response
0: error-free execution of the adjustments
1: indicates that an error occurred; the process has been canceled
:CALibration:ALL[:MEASure]?......................................................................................... 864
:CALibration<hw>:ALL:DATE?.........................................................................................864
:CALibration<hw>:ALL:INFormation?............................................................................... 864
:CALibration<hw>:ALL:TEMP?........................................................................................ 865
CALibration Subsystem