Remote Control Commands
R&S
®
SMW200A
865User Manual 1175.6632.02 ─ 30
Example:
CAL:ALL:INF?
"Instrument is calibrated, no adjustment required."
"UNCAL, instrument is warming up."
"UNCAL, Please perform full adjustment after warming up."
"UNCAL, Please perform full adjustment."
Usage: Query only
Manual operation: See "Information" on page 1281
:CALibration<hw>:ALL:TEMP?
Queries the temperature deviation compared to the calibration temperature.
Return values:
<Temperature> string
Example:
CALibration:ALL:TEMP?
// "+12.00 K"
Usage: Query only
Manual operation: See "Temperature Offset" on page 1281
:CALibration<hw>:ALL:TIME?
Queries the time elapsed since the last full adjustment.
Return values:
<Time> string
Example:
CAL:ALL:TIME?
// "22 days"
Usage: Query only
Manual operation: See "Time" on page 1281
:CALibration:ALL:WARN <UncalWarning>
Generates a warning message, if the instrument is in uncalibrated state.
Parameters:
<UncalWarning> 0 | 1 | OFF | ON
*RST: 0
Example:
CALibration:ALL:WARN 1
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date> string
CALibration Subsystem