Signal Routing and System Configuration
R&S
®
SMW200A
180User Manual 1175.6632.02 ─ 30
Settings:
Signal Output.............................................................................................................. 180
User Variation..............................................................................................................180
Signal Monitoring........................................................................................................ 181
Power-On State...........................................................................................................181
Group Delay Compensation........................................................................................182
Signal Output
Comprises the level settings of the output signal. The signal level is expressed as a
peak or RMS level value.
If R&S SGT is connected to the DIG I/Q interface of the R&S SMW, the level parame-
ters are set automatically.
Note: The "Level" display always refers to both signal components, i.e. SQR(I
2
+Q
2
).
"Set Level Via"
Selects whether the signal level is expressed as a peak or as RMS
level value.
Remote command:
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:POWer:VIA on page 1148
[:SOURce]:IQ:OUTPut:DIGital:FADer<ch>:POWer:VIA on page 1148
"Peak Level"
Enters the peak level of the output signal relative to full scale of 0.5 V
(in terms of dB full scale).
Remote command:
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:POWer:PEP on page 1148
[:SOURce]:IQ:OUTPut:DIGital:FADer<ch>:POWer:PEP on page 1148
"Level"
Enters the RMS level of the output signal.
Remote command:
[:SOURce]:IQ:OUTPut:DIGital:BBMM<ch>:POWer:LEVel on page 1148
[:SOURce]:IQ:OUTPut:DIGital:FADer<ch>:POWer:LEVel on page 1148
User Variation
If the rotary knob is used to adjust the level values, the step width can be defined. The
configured step width applies with activated variation.
"User Variation" is disabled, if R&S SGT is connected to the DIG I/Q interface of the
R&S SMW.