Monitoring Signal Characteristics
R&S
®
SMW200A
631User Manual 1175.6632.02 ─ 30
3. In the task bar, select "System Config > Setup > Baseband Powers" and perform
the following:
a) "Source > Baseband A"
b) "Output > RF A"
c) "Acquisition > Continuous"
d) "Sweep Time = 1 s"
e) "Run Mode > Auto"
f) Select "Add"
g) Select "Start" to trigger the measurement.
The measurement is running automatically with sweep time of 1 s.
The dialog displays the measured values in the units of the RF output signal:
"Level = -30 dBm" and "PEP = -19.06 dBm". These values are identical to the
"Level" and "PEP" values displayed in the status bar.
4. In the block diagram, select "Baseband B > IEEE 802.11" and enable the same
settings as in the first baseband.
5. Select "Baseband B > Signal Routing > Route to Path A and Path B".
Querying Information on the Baseband Power Levels