EasyManua.ls Logo

Rohde & Schwarz RTM3K-COM2

Rohde & Schwarz RTM3K-COM2
861 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Power analysis (option R&S RTM-K31)
R&S
®
RTM3000
343User Manual 1335.9090.02 ─ 09
14 Power analysis (option R&S RTM-K31)
With the R&S RTM3000 and option R&S RTM-K31 you can perform power analysis
measurements.
Probe adjustment..................................................................................................343
Report settings......................................................................................................344
Statistic menu settings.......................................................................................... 345
Input power measurements...................................................................................346
Output power measurements................................................................................359
Switching power measurements........................................................................... 367
Power path power measurements........................................................................ 375
14.1 Probe adjustment
14.1.1 Deskewing the probes
Some power measurements require a current and a voltage probe. Before you start
one of these measurements, you need to deskew the probes to get correct measure-
ment results. The settings menu of the corresponding measurements contain the
"Probe" function to adjust the probes.
Required equipment:
R&S RT-ZF20 power deskew fixture
Rohde & Schwarz voltage probe
Rohde & Schwarz current probe
1. Connect the differential voltage probe and the current probe to the oscilloscope.
2. Press the Apps Selection key. Tap "Power Analysis".
3. Select a measurement.
4. In the configuration menu, select the correct channels for the "Current" and the
"Voltage" sources.
5. Select the "Probe" menu.
6. Connect the probes to the R&S RT-ZF20 power deskew fixture.
7. "Deskew" the probes and adjust the "Zero Offset".
8. Disconnect the probes from the R&S RT-ZF20 power deskew fixture.
Probe adjustment

Table of Contents

Related product manuals