Serial bus analysis
R&S
®
RTM3000
341User Manual 1335.9090.02 ─ 09
SDI ← Label and Data
Sets the values for the source/destination identifier (SDI) bits.
Remote command:
SEARch:PROTocol:ARINc:SDI on page 713
Data ← Label and Data
Searches for a data, that is specified in the submenu.
Data Offset ← Data ← Label and Data
Sets a data offset.
Remote command:
SEARch:PROTocol:ARINc:DATA:OFFSet on page 712
Data Size ← Data ← Label and Data
Sets the data size.
Remote command:
SEARch:PROTocol:ARINc:DATA:SIZE on page 712
Compare ← Data ← Label and Data
Sets the condition how the decoded value is compared with the defined range.
Remote command:
SEARch:PROTocol:ARINc:DATA:CONDition on page 711
Data ← Data ← Label and Data
After setting the "Compare" condition, you can enter the value bit-by-bit by setting the
state high, low, or don't care for each single bit. Alternatively, you can enter a hexadeci-
mal value for each byte.
Remote command:
SEARch:PROTocol:ARINc:DATA:MAXimum on page 711
SEARch:PROTocol:ARINc:DATA:MINimum on page 712
13.9.6 ARINC 429 label list
Label lists are protocol-specific. An ARINC 429 file contains three values for each iden-
tifier:
"Arinc Label": the Arinc 429 label value, that identifies the data type and the param-
eters associated with it.
"Symbolic label": symbolic name of the label, specifying the device function.
"Word Format": an optional parameter to describe the composition of the ARINC
429 word. It can be provided as integer value or mnemonic.
Word format definition
0 or DATA = <P>< Data ><Label>
1 or DATA_SDI = <P>< Data ><SDI><Label>
2 or DATA_SSM = <P><SSM>< Data ><Label>
ARINC 429 (option R&S
RTM-K7)

Table of Contents