Applications
R&S
®
RTM3000
208User Manual 1335.9090.02 ─ 09
Marker value table
There are two markers available for the Bode plot. They are highlighted on the Bode
plot diagram by a white line and the respective marker number 1 or 2. You can move
the markers as needed.
In the marker table you can see the frequency, gain and phase values for both of them
and also the delta values between the two markers.
Remote commands:
●
BPLot:MARKer<m>:DIFFerence:FREQ? on page 567
●
BPLot:MARKer<m>:DIFFerence:GAIN? on page 567
●
BPLot:MARKer<m>:DIFFerence:PHASe? on page 568
●
BPLot:MARKer<m>:FREQuency on page 568
●
BPLot:MARKer<m>:GAIN? on page 568
●
BPLot:MARKer<m>:INDex on page 568
●
BPLot:MARKer<m>:PHASe? on page 568
●
BPLot:MARKer<m>:SSCReen on page 569
Vertical position and size of the waveforms
To set the position and the vertical scaling of the gain, phase or amplitude waveforms,
select the waveform and use the vertical [Scale] and [Offset/Position] (upper knob)
knobs.
Remote commands:
●
BPLot:AMPLitude:SCALe on page 567
●
BPLot:AMPLitude:POSition on page 566
●
BPLot:GAIN:SCALe on page 565
●
BPLot:GAIN:POSition on page 565
●
BPLot:PHASe:SCALe? on page 566
●
BPLot:PHASe:POSition? on page 566
9.8.2 Using a bode plot
To run a Bode plot measurement
1. Prepare the measurement setup. See "Connecting the test setup" on page 209
2. Start the Bode plot option. See "Starting the Bode plot" on page 209
3. Set the required parameters. See "Setting up the Bode plot" on page 209.
Bode plot (option R&S
RTM-K36)