Waveform analysis
R&S
®
RTM3000
142User Manual 1335.9090.02 ─ 09
"Rising"
Only positive clock edges are considered.
"Falling"
Only negative clock edges are considered.
"Either"
The clock edges next to the data edge are considered regardless of
the clock slope. Use this setting, for example, for signals with double
data rate.
Remote command:
SEARch:TRIGger:DATatoclock:CEDGe on page 496
Setup Time
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
Remote command:
SEARch:TRIGger:DATatoclock:STIMe on page 496
Hold Time
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
Remote command:
SEARch:TRIGger:DATatoclock:HTIMe on page 496
7.5.9 Pattern search
For pattern search, up to four analog channels can be used as source. If MSO option
R&S RTM-B1 is installed, also digital channels can be included in the pattern. For each
channel, you define the state. The states are combined logically, and the time of true
pattern results is compared with a specified time range. Thus you can find state transi-
tions inside or outside this time range.
► Select [Search] > "Search Type" = "Pattern" > "Setup".
Search