Remote control commands
R&S
®
RTM3000
464User Manual 1335.9090.02 ─ 09
Parameters:
<State> ON | OFF
*RST: OFF
17.5.3 Edge A/B trigger
TRIGger:B:ENABle.........................................................................................................464
TRIGger:B:SOURce....................................................................................................... 464
TRIGger:B:EDGE:SLOPe............................................................................................... 464
TRIGger:B:MODE.......................................................................................................... 465
TRIGger:B:DELay.......................................................................................................... 465
TRIGger:B:EVENt:COUNt...............................................................................................465
TRIGger:B:ENABle <State>
Activates the B-trigger.
Parameters:
<State> ON | OFF
*RST: OFF
TRIGger:B:SOURce <Source>
Sets the source of the B-trigger.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4 | D0 ... D15
CH1 | CH2 | CH3 | CH4
Analog input channels. Available channels depend on the instru-
ment type.
D0..D15
Digital channels D0 to D15. Require MSO option R&S RTM-B1.
*RST: CH1
TRIGger:B:EDGE:SLOPe <Slope>
Sets the slope of the B-trigger.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive