Remote control commands
R&S
®
RTM3000
795User Manual 1335.9090.02 ─ 09
Parameters:
<DutyCycle> Range: 1 to 99
Increment: 1
*RST: 50
Default unit: %
PGENerator:PATTern:LED:INTens <DutyCycle>
Sets the duty cycle of the signal at P3, which corresponds to the intensity of the light.
You can use this signal if the controller supports separate intensity control.
Parameters:
<DutyCycle> Range: 1 to 99
Increment: 1
*RST: 50
Default unit: %
17.15
Status reporting
17.15.1 STATus:OPERation register
The commands of the STATus:OPERation subsystem control the status reporting
structures of the STATus:OPERation register:
See also:
Chapter B.1, "Structure of a SCPI status register", on page 816
Chapter B.3.3, "STATus:OPERation register", on page 821
The following commands are available:
STATus:OPERation:CONDition?......................................................................................795
STATus:OPERation:ENABle............................................................................................796
STATus:OPERation:NTRansition..................................................................................... 796
STATus:OPERation:PTRansition......................................................................................796
STATus:OPERation[:EVENt]?..........................................................................................796
STATus:OPERation:CONDition?
Returns the of the CONDition part of the operational status register.
Return values:
<Condition> Condition bits in decimal representation. ALIGnment (bit 0) ,
SELFtest (bit 1) , AUToset (bit 2), WTRigger (bit 3).
Range: 1 to 65535
Increment: 1
Usage: Query only
Status reporting