Serial bus analysis
R&S
®
RTM3000
305User Manual 1335.9090.02 ─ 09
Symbolic ID ← Identifier condition
If a label list with symbolic names was loaded and applied in the bus configuration, you
can select a symbolic name from the list instead of entering the numeric identifier. The
instrument triggers on the identifier of the selected node.
Data condition
The data condition consists of the following settings:
Length of the data pattern
Comparison
Data pattern
Data ← Data condition
Defines the length of the data pattern - the number of bytes in the pattern.
Remote command:
TRIGger:A:LIN:DLENgth on page 658
Compare ← Data condition
Sets the comparison condition: If the pattern contains at least one X (don't care), you
can trigger on values equal or not equal to the specified value. If the pattern contains
only 0 and 1, you can also trigger on a range greater than or lower than the specified
value.
Remote command:
TRIGger:A:LIN:DCONdition on page 658
Bin / Hex ← Data condition
Specifies the data pattern to be triggered. To set an individual binary bit or hexadecimal
nibble (half byte), tap it and enter it via the on-screen keypad. Make sure to specify
complete bytes.
"Bin"
String containing the binary pattern with a maximum of 64 bits. Char-
acters 0, 1 and X are allowed.
"Hex"
String containing the hexadecimal pattern with a maximum of 8 bytes.
Characters 0 to F and X are allowed.
Remote command:
TRIGger:A:LIN:DATA on page 657
13.6.4 LIN decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. In the "Bus" menu, enable "Decode".
2. In the "Display" menu, select the result display settings.
See Chapter 13.1.2, "Displaying decode results", on page 253.
3. In the "Bus Table" menu, enable the "Bus Table". Adjust the table settings.
See also: Chapter 13.1.3, "Bus table: decode results", on page 254
LIN (option R&S
RTM-K3)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals