Serial bus analysis
R&S
®
RTM3000
302User Manual 1335.9090.02 ─ 09
Remote command:
BUS<b>:LIN:STANdard on page 655
Bit Rate
Sets the number of transmitted bits per second. The LIN standard defines a maximum
bit rate of 20 kbit/s.
"Predefined"
To select a bit rate from the list of predefined values, set "Bit Rate" to
"Predefined" and select a value from the list.
"User"
To set another value, set "Bit Rate" to "User" and enter a bit/s value.
Remote command:
BUS<b>:LIN:BITRate on page 655
13.6.3 LIN trigger
Before you set up the trigger, make sure that the bus is configured correctly. See
Chapter 13.6.2, "LIN configuration", on page 300.
To trigger on LIN signals:
1. Press the [Protocol] key in the Analysis area of the front panel.
2. Select the bus that is configured for LIN.
3. Select "Trigger".
This selection has several effects:
● Enables decoding, if necessary.
● Sets the "Trigger Type" to "Serial Bus" and the trigger source to the selected
bus.
● Displays the "LIN Trigger" condition in the dialog box, below the protocol setup.
4. At "LIN Trigger", select the required trigger type:
● "Start of Frame": stop bit of the sync field
● "Wake Up": after a wakeup frame
● "Error": checksum, parity or synchronization
● "Identifier": specific message identifier or identifier range
● "Identifier and Data": combination of identifier and data condition
5. If "Identifier" or "Identifier and Data" is selected, the LIN trigger setup dialog
expands to define the serial pattern.
LIN (option R&S RTM-K3)