Remote control commands
R&S
®
RTM3000
434User Manual 1335.9090.02 ─ 09
17.4.1 Automatic setup
AUToscale
Performs an autoset process for analog channels: analyzes the enabled analog chan-
nel signals, and adjusts the horizontal, vertical, and trigger settings to display stable
waveforms
Usage: Event
Asynchronous command
17.4.2 Starting and stopping acquisition
RUN............................................................................................................................. 434
RUNContinous...............................................................................................................434
SINGle..........................................................................................................................434
RUNSingle.................................................................................................................... 434
ACQuire:NSINgle:COUNt................................................................................................434
STOP........................................................................................................................... 435
ACQuire:STATe..............................................................................................................435
RUN
RUNContinous
Starts the continuous acquisition.
Usage:
Event
Asynchronous command
SINGle
RUNSingle
Starts a defined number of acquisitions. The number of acquisitions is set with
ACQuire:NSINgle:COUNt.
Usage:
Event
Asynchronous command
ACQuire:NSINgle:COUNt <NSingleCount>
Sets the number of waveforms acquired with RUNSingle.
Parameters:
<NSingleCount> Number of waveforms.
Range: 1 to maximum number that depends on the record
length.
*RST: 1
Waveform setup