Remote control commands
R&S
®
RTM3000
602User Manual 1335.9090.02 ─ 09
SYSTem:COMMunicate:INTerface:USB:CLASs
Selects the USB mode.
●
USB TMC (Test & Measurement Class)
●
USB VCP (Virtual Com Port)
●
USB MTP (Media Transfer Protocol)
Parameters:
<USBClass> TMC | VCP | MTP
17.10.5
Trigger out
TRIGger:OUT:MODE......................................................................................................602
TRIGger:OUT:PLENgth...................................................................................................602
TRIGger:OUT:POLarity...................................................................................................602
TRIGger:OUT:MODE <OutputMode>
Defines which signals are generated at the [Aux Out] connector.
Parameters:
<OutputMode> OFF | TRIGger | REFerence | MASK
OFF
No output
TRIGger
Outputs a pulse when the instrument triggers.
REFerence
Outputs a 10 MHz reference frequency.
MASK
Outputs a pulse when a mask is violated. This function is only
available if a mask is specified.
*RST: OFF
TRIGger:OUT:PLENgth <PulseLength>
Defines the pulse width of the pulse at the [Aux Out] front connector (at trigger event or
mask viaolation).
Parameters:
<PulseLength> *RST: 1E-6
TRIGger:OUT:POLarity <Polarity>
Defines the polarity of the pulse at the [Aux Out] front connector (at trigger event or
mask viaolation).
General instrument setup