Network connections and remote operation
R&S
®
RTM3000
243User Manual 1335.9090.02 ─ 09
12.2 USB connection
Alternatively to a LAN connection, you can use the USB type B (device USB) connec-
tor at the rear panel to access the instrument via USB.
Connecting the instrument using USB
1. Connect the USB cable to the USB type B connector at the rear panel of the instru-
ment, and to the computer.
2. Open the "Setup" menu.
3. Tap "Interface".
4. Tap "USB" to select USB connection.
5. Tap "Parameter".
6. Select the USB mode.
USB TMC (Test & Measurement Class)
USB VCP (Virtual Com Port)
USB MTP (Media Transfer Protocol)
12.2.1 USB TMC
USB TMC means USB Test & Measurement Class. You can use this interface for
remote control of the instrument using SCPI commands. USB TMC does not need a
driver installation but requires a VISA installation on the controlling computer. VISA is
used to access the instrument, to send remote commands, and to read status informa-
tion.
Rohde & Schwarz provides the standardized I/O software library R&S VISA for com-
munication via TCP/IP (LAN: HiSlip, VXI-11) or USB (USBTMC) interfaces. R&S VISA
is available for download at the Rohde & Schwarz website www.rohde-schwarz.com/
rsvisa.
USB connection