Waveform analysis
R&S
®
RTM3000
139User Manual 1335.9090.02 ─ 09
Remote command:
SEARch:TRIGger:RISetime:RANGe on page 493
Rise/Fall Time
Sets the reference rise or fall time, the nominal value for comparisons.
Remote command:
SEARch:TRIGger:RISetime:TIME on page 493
Variation
Sets a range Δt to the reference "Rise/Fall Time" if comparison is set to "Equal" or "Not
equal". The instrument finds rise/fall times inside or outside the range width ± Δt.
Remote command:
SEARch:TRIGger:RISetime:DELTa on page 493
7.5.7 Runt setup
The runt search finds pulses lower than normal in amplitude. The amplitude crosses
the first threshold twice without crossing the second one. In addition to the threshold
amplitudes, you can define a time limit for the runt in the same way as for width search:
runts with exact width, shorter or longer than a given time, or runts inside or outside the
allowable time range.
[Search] > "Search Type" = "Runt" > "Setup"
Polarity
Indicates the polarity of the pulse to be searched for.
Remote command:
SEARch:TRIGger:RUNT:POLarity on page 494
Search

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals